I own 4 versions of the NanoNVA and have found it incredibly useful. However, I never stopped to question just how accurate or reliable the measurements are. That was until Matthew, VK5ZM, was in my workshop on the weekend and he made a suggestion in response to some concerns I was having over the measurement of attenuators. The outcome of following that suggestion was a better understanding of just how good the NanoVNA is and what might be limiting it's performance.
The assumption I implicitly made was the SOLT calibration process reduced errors in measurement to an immaterial level. Rather than lots of mathematics to validate or invalidate the NanoVNA you should repeat the following steps with your instrument to decide for yourself what you are prepared to accept as correct, close enough or simply wrong.
Ideally you will have a SMA attenuator, 10-15dB in value, that you can insert before the 50 ohm calibration load.
Step1
Perform a SOLT calibration directly on the SMA port, with the load being the attenuator / reference load combination and the Through step using one of the supplied cables. I used the NanoVNA app to average 8 readings and a 1 to 801MHz range with 401 or 801 steps.
Step 2
Now reattach the attenuator / reference combination and measure the input return loss. Save that to perhaps M1.
Step 3
Now remove the reference load leaving just the attenuator attached to the measurement port. You should see a chart like this:
The purple trace is the input return loss of the load plus attenuator combination and two things stand out: Noise ( even after averaging ) and the step change in measurements once the frequency was higher than 300MHz.
The red line is the attenuator alone and as expected the return loss was twice the attenuation (2x-10dB).
That red line confirms that input return loss measurements at the NanoVNA SMA connector down to -20dB can be accepted as correct, anything down to -45dB up to 300MHZ and perhaps -35db to -30dB above that is probably close enough. And any return loss measurements approaching the purple line, or below that, is probably invalid and you should be rounded up to whatever "close enough" reference you decide upon.
Step 4.
Save your calibration and then attach the attenuator / reference load combination at the point you normally calibrate to. For me that's at the end of a SMA to SMA cable some 25cm in length. It could be one of the supplied cables but I replaced mine since one became damaged.
Repeat the SOLT calibration at the end of the SMA cable (Reference plane) then measure the input return loss of the attenuator load combination, and the attenuator by itself.
My results are shown below:
The red line, attenuator only, is essentially unchanged but the attenuator / load combination had shifted upwards and noise has reduced.
At this point you should be starting to question why this is so. Is the cable lossy? Is this a consequence of the SOLT calibration? Why did this happen?
Step 5 - Be afraid. This will change your life forever.
OK, it's not that bad but now you will see some results that should forever change your opinion on how good your measurements are.
Restore the calibration to the values saved when the attenuator / load were directly attached to the NanoVNA. Attach your SMA cable to the NanoVNA and the attenuator / load on the far end. Blink in disbelief and sit down. A 50 ohm load at the end of a 50 ohm cable should measure the same return loss as when the cable was not present. If the cable had loss then the return loss would be lower.
Blue SMA cable supplied with NanoVNA V4 in metal box (ie the supposedly top end version)
Clearly it's not a 50 ohm cable. Calibration means I never knew this
cable was rubbish since calibration hides how bad this cable is. But it puts into question past
measurements I made using this cable with the "Top End" NanoVNA model.
The black SMA to SMA cable supplied with the lower end NanoVNA is slightly better but still rubbish:
And now a collection of results as I tried to look for a good cable.
|
Long Cable SMA one end, N Connector the other
|
|
Hardline - SMA to N |
|
|
|
Double shielded Silver plated - SMA to N (A quality cable) |
|
SMA to N Pigtail from AliExpress
|
Conclusion
I now appreciate why high end equipment comes with very expensive
connecting cables. The intrinsic ability of the NanoVNA to measure input
return loss is quite good, especially below 300MHz. However, the cables
used can result in meaningless measurements below a certain threshold
despite the SOLT calibration. I seems I don't own a single cable with SMA connectors each end that I would accept for measurement purposes.
While the SOLT calibration can reduce the influence of cables and connectors up to the reference plane there will be a point where measurement uncertainty creeps. And if you flex, twist, curl or bend the cable between the reference plane and NanoVNA the cable impedance could, probably will, change which again introduces measurement uncertainty.
Measurement uncertainty means that when you are tuning / matching a load to 50 ohms is -35dB actually better than -30dB? Is a -20dB return loss the influence of cables/connectors or is it a true measurement of the load? I have noted from past testing with a homebrew return loss bridge that different examples of even the best quality 50 ohm cables I have terminated in N Connectors can show several dB of difference when measuring return losses.
If you want to reduce your measurement uncertainty you will need to:
- buy/make better connecting cables,
- test them with an attenuator in the fashion described.
Only then should you be comfortable relying on the measurements.
Whether you call it VSWR, R+jX, a smith chart or input return loss
measuring reflections from the load is one of the key uses of the
NanoVNA. It is well worth the few minutes it takes to test the cables so you understand when your measurements become uncertain.
I'm off to buy a -15dB SMA attenuator to improve my comfort level in the measurements being made. I'll post an update once I have procured one and repeated the tests.